文件编号:NR-TP-0715-008
2.2

Environmental Requirements

The storage, transport and use of the monitor must meet the following environmental
requirements.
Working
environment
Supply voltage
requirement
Transportation
Storage
The operating environment of the monitor should avoid noise, vibration, dust, corrosive or
flammable and explosive materials. In order to allow air flowing smoothly and achieve good
heat dissipation, at least 2 inches (5cm) clearance should be kept around the device.
When the device is moved from one environment to another, the device may have
condensation due to differences in temperature or humidity. In this case, wait until the
condensation disappears before using the device.
Warning
Please ensure that the device operates under stipulated environment, or else
technical specifications declared in this Manual may not be met, and it may
result in damage to equipment and other unforeseen consequences.
2.3

Connecting to AC Power

Connect to AC power in the following steps:
Ambient
temperature range
Relative
humidity
range
Atmospheric
pressure range
Supply voltage
Power frequency
Input power
Fuse
Prevent severe shock, vibration, rain and snow splashing during transport.
The packaged monitor should be stored in well-ventilated room with
ambient temperature -20℃~+40℃, relative humidity lower than 80%,
atmospheric pressure 50kPa~106kPa, and without corrosive gases.
5℃~40℃
≤80%
86kPa~106kPa
a.c.100V~240V
50Hz/60Hz
40VA-60VA
FUSE T 1.6A
版本: 1.0
- 6 -