文件编号:NR-TP-0715-008
C.2 Technical Alarm Information
Source
Alarm message
System Battery Low
ECG
ECG Comm. Stop
ECG Comm. Error
ECG Config Error
ECG Self check Error
ECG Lead Off
ECG YY OFF (YY is
a lead name)
TEMP1 Sensor Off
Temp
TEMP2 Sensor Off
SPO2 Comm. Stop
SPO2 Comm. Error
SPO2 No Sensor
SPO2 Sensor Off
SPO2 Sensor Error
SpO2
SPO2 Search Timeout
SPO2 Search Pulse
SPO2 Signal Unstable
SPO2 Failure
SPO2 Signal Weak
NIBP Comm. Stop
NIBP Comm. Error
NIBP Self check error
NIBP
NIBP CFG Error
NIBP system error
Measurement timeout
Causes and countermeasures
L
Connect to AC power supply, and charge the
H
battery, and power with the battery as needed after
fully charged.
H
ECG module failure, or communication failure
between the module and the host; please restart
H
the device.
H
H
M*
The electrodes are not connected to the patient
firmly or fall off, or lead wires and the main cable
fall off. Check the connection of electrodes and
M*
lead wires.
L
The temperature sensor falls off from the patient.
Check the sensor connection.
L
H
SpO2 module failure, or communication failure
between the module and the host; please restart
H
the device.
L
SpO2 sensor falls off from the patient or monitor,
malfunctions, or sensor other than specified in this
L
Manual is used. Check the sensor mounting
L
position, whether the sensor is damaged or sensor
L
type. Reconnect the sensor or use new sensor.
L
Sensor signal is poor or too weak. Check the
patient's condition, and place the sensor in a
L
suitable position. If the failure persists, replace the
L
sensor.
L
H
NIBP module failure, or communication failure
between the module and the host; please restart
L
the device.
H
H
H
If failure occurs during measurement, the system
can't analyze and calculate. Check the patient's
L
版本: 1.0
- 93 -