文件编号:NR-TP-0715-008
Appendix A Specifications
A.1 Safety Specifications
A.1.1
Product Category
In accordance with classification specified in the European Medical Device Directive
93/42/EEC, this monitor is IIb type equipment. The monitor is classified as follows in
accordance with IEC 60601-1:
Category Name
93/42/EEC ohm
Medical Devices
Directive Category
Type of electric
shock protection
Electric shock
protection grade
Explosion
protection grade
Liquid inlet
protection grade
Operating mode
Movement
A.1.2
Power Specifications
Parameter
External AC power
Input voltage
Input current
Frequency
Internal power supply: Lithium-ion battery
Specification
IIb
Grade I, internal and external power supply equipment
When you question the integrity of the external protective earthing or
protective ground conductor parameter of the equipment, the device must
be powered by the internal power supply (battery).
ECG/NIBP/SpO2/PR/TEMP/RESP: CF (defibrillation protection)
Common equipment, no explosion protection
IPX1 (prevent water from entering when the water drips vertically)
Continuous
Portable
Specification
100-240V~
1.5A
50/60Hz
版本: 1.0