Inflation Pressure; Resetting; Leakage Test - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008

10.6 Inflation Pressure

If necessary, you can manually set the initial cuff inflation pressure as follows:
Select the NIBP parameter area → [NIBP Setup] menu;
Select [Pre-Infl Press], and set the appropriate cuff pressure value.

10.7 Resetting

Select NIBP parameter area → [NIBP Setup] menu → Select [Reset], and restore the
inflation pressure of the blood pressure pump to currently configured initial settings. When
the blood pressure pump is not working properly, but no warning is given, you can reset the
blood pressure pump, and automatically restores the blood pressure pump.

10.8 Leakage Test

The purpose of leakage test is to detect if the sealing of the air passage is in good condition.
If the leakage test passes, the NIBP parameter area displays [Leakage test Stopped]. If not
passed, the NIBP parameter area displays [Cuff leak] message. NIBP leakage test shall be at
least once every two years or when you think that the reading is not accurate.
Prepare the following materials before the test:
Adult cuff: one
Inflation tube: one
Cylinder of appropriate size: one
Leakage test process
Fig. 11-1 NIBP Parameter Area
版本: 1.0
- 52 -