Chapter11 Temp; Overview; Safety Information; Measurement Steps - Northern Meditec Taurus E Operator's Manual

Multi-parameter patient monitor
Table of Contents

Advertisement

文件编号:NR-TP-0715-008

11.1 Overview

This monitor has two temperature measurement channels; the temperature sensor will
measure the body temperature, and calculate the difference between the body temperature
data.

11.2 Safety Information

Warning
Before monitoring, check if the probe cable is normal. Unplug the temperature
probe cable from the jack, the screen will display [TEMP1/TEMP2 Sensor Off]
prompt and make an alarm sound.
Calibrate the temperature measuring instrument at least once every two years (or
according to hospital procedures). When calibration is required, please contact the
manufacturer.

11.3 Measurement Steps

Please refer to the following steps:
1. Turn on the monitor and check if it works normally.
2. Select the appropriate temperature probe according to the patient category and
measurement needs.
3. Insert the probe lead wire into the temperature probe interface.
4. Attach the probe to the patient properly.
5. Make sure that the alarm settings apply to the patient.

11.4 Measuring Requirements

The normal measuring range of body temperature is 5~50°C, and the accuracy is consistent
in this range.

Chapter11 Temp

版本: 1.0
- 55 -