KaVo DIAGNOcam Vision Full HD Instructions For Use Manual page 32

Table of Contents

Advertisement

Instructions for use DIAGNOcam Vision Full HD
6 Operation | 6.4 Recording images
▪ Metabolic products of cariogenic bacteria are shown in a reddish colour
Recording an image in Fluorescence mode
▶ Connect the camera to the computer.
▶ Start up the imaging software.
▶ Remove the camera from the camera holder.
▶ Before the first use process the Fluorescence Cover in accordance with the
instructions for use.
See also:
2 7 Reprocessing steps in accordance with ISO 17664, Page 37
▶ Apply the sheath first before you attach the Fluorescence Cover to the cam-
era.
See also:
2 6.2 Use a sheath, Page 26
To be able to capture an optimum image, the Fluorescence Cover needs to be
fully plugged onto the camera. This prevents the ingress of extraneous light
and is an additional aid for simple positioning of the camera on the tooth.
The recording window of the imaging software shows a live image.
▶ Select Fluorescence mode (still image or video) in the imaging software or
with the menu button.
▶ Select an image detail.
▶ The camera automatically focuses the selected image detail (permanent
autofocus).
▶ Press one of the two trigger buttons.
▶ The camera switches to Freeze mode and/or starts the video recording. The
still image/video is transferred to the imaging software.
▶ To return to Live mode and/or to stop the video recording, press the trigger
button again.
▶ Process and save the image/video in the imaging software.
32 / 52

Advertisement

Table of Contents
loading