KROHNE OPTIWAVE-M 7400 C Short Manual page 63

24 ghz radar (fmcw) level transmitter foragitated and corrosive liquids
Table of Contents

Advertisement

OPTIWAVE-M 7400 C
Screen
02/2019 - MA OPTIWAVE-M 7400 C R01 en
Steps
[ ] or [ ] to change the parameter
("No" or "Yes").
[^] to confirm.
[^] to confirm.
9
[ ] or [ ] to change the parameter
("No" or "Yes").
[^] to confirm.
[ ] or [ ] to change the parameter
("Disable" or "Enable").
[^] to confirm.
4 × [^] to confirm.
[ ] or [ ] for the selection of the
save option (Yes, No or Back).
[^] to confirm.
www.krohne.com
OPERATION
Description
Start recording? Set this menu item to "Yes" to
continue to the subsequent step. Set this menu
item to "No" to go back to the menu.
Graph with the results of the empty spectrum
recording.
Save spectrum? If you set the menu item to
"Yes", the device will keep this empty spectrum
recording. If you set the menu item to "No", the
device will reject this data.
Empty Spectrum Enable. If you set the menu
item to "Enable", the device will use the empty
spectrum recording data. If you set the menu
item to "Disable", the device will not use the
empty recording data at this time.
Save Configuration? screen.
Set to "Yes" to save and use the settings and
data and go back to Normal mode. Set to "No"
to cancel the changes to the device settings and
go back to Normal mode. Set to "Back" to stay in
Program mode.
4