02/2019 - MA OPTIWAVE-M 7400 C R01 en
Function description
This sets the device to a given test level
value (linearized). This menu item is only
available if you set up a linearization table
in menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Lin. Level" before the device
simulates a level value. This procedure
sends an output signal that agrees with the
test reading. Output will change to the
selected value, independent of the
measurement data.
When you push [^] to confirm the value,
the display shows the question "Start
Simulation?" Push [ ] or [ ] to set the
display to "Yes". Push [^] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
This sets the device to a given test volume
value. This menu item is only available if
you set up a conversion (volume) table in
menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Volume" before the device
simulates a volume value. This procedure
sends an output signal that agrees with
the test reading. Output will change to
the selected value, independent of the
measurement data.
When you push [^] to confirm the value,
the display shows the question "Start
Simulation?" Push [ ] or [ ] to set the
display to "Yes". Push [^] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
This sets the device to a given test mass
value. This menu item is only available if
you set up a conversion (mass) table in
menu item C3.1 Edit Table. Make sure
that menu item C4.1.1 Current Out. 1 Var.
is set to "Mass" before the device
simulates a mass value. This procedure
sends an output signal that agrees with
the test reading. Output will change to
the selected value, independent of the
measurement data.
When you push [^] to confirm the value,
the display shows the question "Start
Simulation?" Push [ ] or [ ] to set the
display to "Yes". Push [^] again to start
the test. The device goes to back to
Normal mode after 1 hour.
Minimum access level to do the
procedure: Expert
www.krohne.com
OPERATION
Selection list
min-max:
-5000.0...+5000.0 m /
+03
+03
-196.85
...+196.85
in /
-16404...+16404 ft
min-max:
0
+06
.... 1.00
min-max:
0
+09
.... 10.000
kg
4
Marine
Default
1
1
1
41