Current output and LCD display settings
When you do the Standard setup procedure, the measurement value on the first measurement page of
the LCD display (in Normal mode) automatically uses settings with the same parameters and values as
the current output settings (measurement variable, 0% range and 100% range). The first measurement
page shows "one value and bar" by default. If you must show one or more measurement values that are
different to the transmitted output signal, change the settings in the C6.4 1st Meas. Page and C6.5
2nd Meas. Page menus. For more data, refer to Function description on page 38
menu (C6 Display).
Measurement data includes different measurement types (level, distance, volume, ullage volume, mass
etc.). Some measurement types will only be available in Normal mode if you entered the correct
parameters in Program mode. Normal mode shows abbreviations of measurement types if the
measurement page is configured to show more than one measurement value. For a list of the
abbreviations used in Normal mode, refer to the table that follows:
Abbreviations of measurement types used in Normal mode
Measurement type
Distance
Level
Reflection
Sensor value
Volume
Ullage volume
Mass
Ullage mass
Linearized distance
Linearized level
Measurement data format errors
Figure 4-2: Error symbol: the number of digits and decimal places is not sufficient for the measurement data
1 Error symbol: the number of digits and decimal places is not sufficient for the measurement data. It is possible that the length
units must be changed from "mm" to "m".
In this example, the device measures a distance of 10.001 m, but C7.5.1 Length is set to "mm" and C6.4.5
Format 1st Value is set to "X.XXX" (four digits with three decimal places). This is not sufficient to show a
measurement value of 10.001 m. If the measurement value is equal to or more than 10 m, set C6.4.5
Format 1st Value to "Automatic".
You can change the number of digits and decimal places in the measurement values shown in Normal mode.