PROBLEM
Unit does not Power On
Voltage alarm condition
(RED LED illuminated).
Temperature alarm
condition (RED LED
illuminated).
Fan alarm condition
Maintenance 5-12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table 5-1, Troubleshooting

POSSIBLE CAUSES
1264C mainframe not
connected to power
source.
Blown AC Mains Fuse.
Power supply protections
are active causing the
supply to be "shutdown".
Faulty AC wiring or bad
front panel AC Power
Switch.
Faulty VXI module
installed or voltage fault
occurred.
VXI supply is out of
tolerance.
Supply intake
temperature limit
exceeded.
or
Insufficient module
cooling air.
Fan intake is blocked, or
fan filter is clogged.
S400 Voltage switch set
to incorrect position.
1264C User Manual
WHAT TO DO
Make sure that the 1264C is
connected to a live electrical outlet and
the AC power switch is on. Try
operating another piece of equipment
from this outlet.
Unplug AC mains power and check the
fuse. See "Removal and Replacement
of the AC Mains Fuse" in Chapter 5.
Refer to "Power Supply Protections" in
Chapter 4.
Cycle power to clear fault. If fault
persists remove installed VXI modules
(cycle power) until fault is cleared.
Verify integrity of AC wiring to AC
Power switch and the switch itself.
Contact customer service.
Cycle power.
If fault persists remove installed VXI
modules until fault indication is
cleared.
Perform Basic Functional Check
procedure from Chapter 3.
Check for restrictions to airflow at
mainframe intake and exhaust.
Clean mainframe module cooling fan
filter.
Check ambient temperature.
Check for restrictions to airflow at
mainframe intake and exhaust.
Clean mainframe module cooling fan
filter.
Set S400 to position that corresponds
to the input power voltage.