Power Supply Protections; Backplane - Eads Racal Instruments 1264C Manual

6-slot vxibus mainframe
Table of Contents

Advertisement

1264C User Manual
Power Supply
Protections

Backplane

EADS North America Defense
Test and Services, Inc. © 1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The seven VXIbus power supply outputs are protected for:
Over-voltage
Over-current and short circuit
Over-temperature
The AC input of the power supply includes protections for:
Input current fusing
See Appendix A for specification details pertaining to the above
protections.
If any of the over-voltage or over-temperature protections are
active, the power supplies will shutdown. This means the supply
has turned itself off to protect the system. The AC Power switch
S100 must be cycled to clear the shutdown condition. If the supply
continues to shutdown then the external fault is still present.
If any of the over-current and short-circuit protections are active
the power supply will go into current-limit and the voltage on the
respective output may be reduced or at ground potential. These
protections are self-correcting and the respective output will return
to its nominal value when the fault is removed.
The Backplane (405130) serves several functions
Accommodates up to 6 C-size VXI.
Rigid mechanical interface which accommodates a lifetime
of insertions of VXI modules.
Mechanically connects (at E1 to E14) to the power supply
and distributes DC voltages and currents to VXI modules
(through P1 and P2).
Connects the VME communications interface across P1
and P2 to all slots.
Connects the VXI extension signals across P2 rows A and
C to all slots.
Connects to the system monitor board through a 14-pin
ribbon cable between J50 on the backplane and J301 on
the power supply assembly (407645).
Understanding The 1264C 4-3