Monitoring Basics; Table 3-1, Monitoring Basics - Eads Racal Instruments 1264C Manual

6-slot vxibus mainframe
Table of Contents

Advertisement

1264C User Manual
Monitoring
Basics
WHAT YOU SEE
ON THE FRONT PANEL
Green LED illuminated for:
+5V
+12V
+24V
-2V
-5.2V
-12V
-24V
Green LED illuminated for:
FAN
Green LED illuminated for:
TEMP
EADS North America Defense
Test and Services, Inc. © 1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
4. Observe that the front panel LED indicators all turn Green.
Note that the FAN LED indicator will illuminate RED while
the fan ramps up to full speed. After approximately 20
seconds the FAN LED should turn Green.
Option 71 (P/N 407676) is required for applications powered
by 230 VAC power. The 1264C is configured for 230 VAC
operation ONLY when Option 71 is ordered.
Refer to Figures 3-1 and 3-2 while using Table 3-1 to interpret
system monitor information. Refer to Section 7 Troubleshooting if a
system fault is detected.

Table 3-1, Monitoring Basics

VXI Voltages are within acceptable limits
+4.56V ≤ +5V Rail ≤ +5.50V
+10.95V ≤ +12V Rail ≤ +13.20V
+21.90V ≤ +24V Rail ≤ +26.40V
-2.20V
-5.72V
-13.20V ≤ -12V Rail ≤ -10.95V
-26.40V ≤ -24V Rail ≤ -21.90V
Cooling fan is operating
Power supply intake air temperature ≤ 60°C
WARNING:
WHAT IT MEANS
-2V Rail ≤ -1.83V
≤ -5.2V Rail ≤ -4.75V
Operating The 1264C 3-3