Basic Functional Check; Table 3-5, Power Supply Voltages At The P1 And P2 Connectors - Eads Racal Instruments 1264C Manual

6-slot vxibus mainframe
Table of Contents

Advertisement

1264C User Manual
Basic Functional
Check

Table 3-5, Power Supply Voltages at the P1 and P2 Connectors

Measurement
Location (PIN)
P1-C9, P2-C3
EADS North America Defense
Test and Services, Inc. © 1998
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The functional check consists of checking the 1264C power supply
voltages to the VXIbus limits at the P1 and P2 connectors using a
digital voltmeter.
Referring to Tables 3-2 to 3-5 and Figure 3-1 connect one lead of
the voltmeter to a supply pin. Connect the reference lead of the
voltmeter to one of the ground pins.
reading to the values listed in Tables 3-5.
P1/P2
Supply
P1-C32
+5V
P1-A31
-12V
P2-C32
-24V
P2-C13
-2V
P1-C31
+12V
P2-C31
+24V
P2-C4
-5.2V
Logic Ground
If the voltages are within the specified ranges, the mainframe
complies with the VXI voltage limit specifications.
Note that the "Voltage Range" values in Table 3-5 are identical to
the VXIbus Specification. The voltage range limits provided in
Table 3-1 refer to the voltage monitor indicators on the 1264C front
panel. The monitor will detect a voltage error when the limit is
exceeded (V
nominal
VXI Acceptable
Voltage Range
4.875V to 5.25V
-12.6 to 11.64V
-25.2V to -23.28V
-2.1V to -1.9V
11.64V to 12.6V
23.28V to 25.2V
-5.46V to -5.044V
± 10% V
± 4% accuracy).
nominal
Operating The 1264C 3-7
Compare each voltage