To change a parameter of a DPOAE or TEOAE Test Protocol proceed as follows:
CHANGE
CHANGE
SETUP
CHANGE
(press for 3 s)
NEXT
(press for 3 s)
NEXT
CHANGE
DONE
NOTE: If you push the DOWN arrow key without holding it for 3 seconds, you will
scroll through date and time, etc., rather than accessing the custom menus.
5.10.3.2 Advanced Options for DPOAE Testing
Selecting the Level of Primary Tones
Figure 69
1
8106556 Rev. 11
Operation Manual ERO•SCAN
When a change is made within the Advanced Options menu to
a default protocol, an * is place in the protocol name (Figure 68).
to enter the PROTOCOL menu.
to select the DPOAE/TEOAE protocol (DP 4s and TE 64s
protocols are not customizable).
to enter the Clock menu.
to enter the Setup menu.
Press and hold for 3 s until the green LED (READY) turns off.
to enter the DPOAE/TEOAE menu.
Press and hold for 3 s until the green LED (READY) turns off.
to select a protocol parameter to change.
to change a parameter.
to save the parameters and exit the TEOAE menu.
In the LEVEL menus (LEVEL L1, LEVEL L2) you can set the
level for the primary tones (L1, L2) (Figure 69).
CHANGE to set the level L1 to a value between
40 dB SPL and 70 dB SPL (1 dB steps).
NEXT
to proceed to the LEVEL L2 menu.
CHANGE to set the level L2 to a value between
40 dB SPL and 70 dB SPL (1 dB steps).
NEXT
to proceed to the next screen.