Maico ERO-SCAN Operation Manual page 57

Screening and diagnostic version
Hide thumbs Also See for ERO-SCAN:
Table of Contents

Advertisement

Figure 68
To change a parameter of a DPOAE or TEOAE Test Protocol proceed as follows:
CHANGE
CHANGE
SETUP
CHANGE
(press for 3 s)
NEXT
(press for 3 s)
NEXT
CHANGE
DONE
NOTE: If you push the DOWN arrow key without holding it for 3 seconds, you will
scroll through date and time, etc., rather than accessing the custom menus.
5.10.3.2 Advanced Options for DPOAE Testing
Selecting the Level of Primary Tones
Figure 69
1
8106556 Rev. 11
Operation Manual ERO•SCAN
When a change is made within the Advanced Options menu to
a default protocol, an * is place in the protocol name (Figure 68).
to enter the PROTOCOL menu.
to select the DPOAE/TEOAE protocol (DP 4s and TE 64s
protocols are not customizable).
to enter the Clock menu.
to enter the Setup menu.
Press and hold for 3 s until the green LED (READY) turns off.
to enter the DPOAE/TEOAE menu.
Press and hold for 3 s until the green LED (READY) turns off.
to select a protocol parameter to change.
to change a parameter.
to save the parameters and exit the TEOAE menu.
In the LEVEL menus (LEVEL L1, LEVEL L2) you can set the
level for the primary tones (L1, L2) (Figure 69).
CHANGE to set the level L1 to a value between
40 dB SPL and 70 dB SPL (1 dB steps).
NEXT
to proceed to the LEVEL L2 menu.
CHANGE to set the level L2 to a value between
40 dB SPL and 70 dB SPL (1 dB steps).
NEXT
to proceed to the next screen.
55
®
13//08/2020

Advertisement

Table of Contents
loading

Table of Contents