5.10.2.6 Minimum Value (MIN VALUE)
Figure 60
5.10.2.7 Clock Mode
Figure 61
5.10.2.8 Graph Style
Figure 62
1
8106556 Rev. 11
Operation Manual ERO•SCAN
With the MIN VALUE setting (Figure 60) the user is allowed to
include the minimum OAE response amplitude to the
PASS/REFER criterion for a single frequency. When the MIN
VALUE setting is set between -10 dB SPL and + 5 dB SPL the
OAE response for the frequency under test needs to be at an
equal or higher level than the MIN VALUE set to get a PASS
result. If MIN VALUE is set to OFF, the PASS decision is based
on the SNR alone for the current frequency measured.
Default MIN VALUE setting is -10 dB SPL.
CHANGE
to select a value between -10 dB SPL to 5 dB
SPL or to set it OFF.
NEXT
to proceed to the next screen.
The CLOCK MODE menu allows the user to change the clock
from a 24 hour mode to a 12 hour mode (Figure 61).
CHANGE
to select 12 Hour or 24 Hour.
NEXT
to proceed to the next screen.
The user can select between two options of displaying the result.
(Figure 62). The SNR graph displays the Signal-to-noise ratio as
a function of frequency. The value graph view shows absolute
OAE levels and noise levels as a function of frequency.
Section 5.7.1 Calibration and Test
CHANGE to select GRAPH SNR or GRAPH VALUE.
NEXT
to proceed to the next screen.