5.10.2.13 Reset to Default
Figure 67
5.10.3 Advanced Settings Menus (Diagnostic Version only)
5.10.3.1 Customizing Advanced Settings
The Advanced Options menu permits modification of the test parameters and pass
criterion for the customizable protocols. Changes to the protocol should be made only
by qualified personnel, usually the administrator. If you are not familiar with the use of
these variables, do not attempt to change the protocols. Changes to any of these
characteristics may yield test results that differ from those obtained in other test modes.
IMPORTANT NOTE: Any change of the parameters Level, SNR for PASS, Number of
Frequencies for PASS and Minimum Value has an influence on the Sensitivity and
Specificity of the PASS an REFER decision. If the criterion for PASS is set too loose,
there is a risk, that the test result is a PASS even if a hearing loss is present. If you
choose other settings that the default values, you are responsible.
Section 5.10.3
Section 5.10.2.6
The ERO•SCAN
®
changes are saved in the non-volatile memory so the settings will be retained even
when the battery is discharged temporarily.
Section 6.5
1
8106556 Rev. 11
Operation Manual ERO•SCAN
In the RESET TO DEFAULT menu, you can reset the device
settings to the factory settings (Figure 67).
NOTE: By resetting the device settings, you delete all existing
test results and reset all system and protocol settings. This
procedure also separates the ERO•SCAN
wireless devices.
RESET
YES or NO
NEXT
Advanced Settings Menus (Diagnostic Version only)
Minimum Value (MIN VALUE)
comes with pre-programmed protocol settings. Test protocol
Configurations and Test Protocols
to reset the device settings to default.
to confirm or stop the reset.
to proceed to the next screen.