5.8.2.2 Understanding the Thermal Printout (TEOAE)
The following information is provided for each test (Figure 41 – TEOAE printout with
Value Graph view, Figure 42 – TEOAE Printout with SNR Graph view.
Section 5.10.2.8
Figure 41
Figure 42
1
8106556 Rev. 11
Operation Manual ERO•SCAN
Graph Style
1. Manufacturer Logo
2. The software version number (e.g.: V105.05)
3. The time and date of the test, based on the setting of
the internal clock; if the clock is set correctly, this time
and date will be correct
4. The test number (if operating in Save 500 mode)
(e.g.:001)
5. The protocol selected (e.g.: TE 64s)
6. The test time used to complete the test (e.g.: 4 sec
avg.)
7. Device/Probe serial number (SN)
8. The ear selected (Right or Left)
9. A PASS/REFER indication if there is a criterion set for
the selected protocol
10. Graphic display of results (SNR or Value graph)
11. Printout notification when Minimum Amplitude is 'on'
12. The frequency band center (F)
13. SPL peak equivalent of presented click (L)
14. The level of the emission in dB SPL (TE)
15. The noise floor in dB SPL (NF)
16. The signal-to-noise ratio (SNR) = TE – NF
17. A "P" indicates that the frequency passed based on the
criterion settings within the selected protocol