5.10.2.4 Auto Shutdown Time
Figure 57
5.10.2.5 Save Mode/Storing Test Results
Figure 58
Figure 59
1
8106556 Rev. 11
Operation Manual ERO•SCAN
The POWER OFF Menu allows you to set a time after which the
device shall switch off automatically to prolong battery life
(Figure 57).
NOTE: The device will immediately power off after printing
regardless of this setting.
CHANGE
to select a value between 0.5 and 4 min.
NEXT
to proceed to the next screen.
The ERO•SCAN
®
can store either the most recent test results for
the right and left ear (Figure 58) or the last 500 tests (Figure 59).
CHANGE to select SAVE L/R TESTS or SAVE 500
TESTS.
NEXT
to proceed to the next screen.
When using the 500 Tests mode with the ERO•SCAN
are two ways of operating.
When a Software is used to transfer patient names to the
device, 50 patients can be transferred to the device. The
ERO•SCAN
®
will show the names in an alphabetical order.
You can cycle through the names using the left and right
arrow. A patient named "Unnamed" is always included at the
beginning of the ERO•SCAN
is being tested but the patient name was not transferred to the
ERO•SCAN
®
.
When no patient names are transferred to the device, the
ERO•SCAN
®
will automatically number each test from 1 to 500.
NOTE: When 495 tests have been saved, the user will be
warned that the memory is almost full. When the ERO•SCAN
device reaches 500 saved tests, it will not allow any further
testing. At this point either the results must be printed,
transferred to the PC software, or they must be cleared from
memory.