Maico ERO-SCAN Operation Manual page 21

Screening and diagnostic version
Hide thumbs Also See for ERO-SCAN:
Table of Contents

Advertisement

Table 2 Troubleshooting
Problem
Device does not
turn on
Test does not
start
Results are not
printed
Display is frozen
and device does
not respond to
button presses
Numerical
results show a
line rather than a
number
1
8106556 Rev. 11
Operation Manual ERO•SCAN
Explanation
• Press the DOWN arrow for a full second (the yellow
LED (TEST) illuminates).
• Connect the charger as shown in Figure 2 on page
37. Confirm that the blue LED (CHARGE) is
illuminating in a slow blink pattern. Wait at least 10
minutes and then attempt to turn on the device.
• Select a different sized eartip.
• Reposition the probe.
• Change the probe tube.
• Verify that the eartip is sealed in the ear canal via
feedback from the PROBE CHECK screen.
• Try if the device starts in your own ear with the proper
eartip. If the test does not start or if the AutoStart
tones sound unusual, replace the probe tube.
• Check the printer status. Turn the printer on (wake
from sleep mode) by pressing the large button.
• If the printer does not turn on, plug in the power supply
to charge the battery.
• Make sure the printer has paper.
• If paper comes out of the printer, but nothing is printed
on the paper, the paper roll may have been inserted
the wrong way round.
• Press the large printer button twice rapidly to run
demo print.
• Press and hold the DOWN arrow button for 10
seconds to force the device to power off. The device
should function normally the next time it is switched on.
Dashes are displayed when there is excessive noise while testing.
• Reduce the noise and retest.
Dashes are displayed in the DP and SNR column when the
signal level is at a minimum of 5 dB over target level.
• Reinsert the probe and retest.
19
®
13//08/2020