Figure 30
NOTE: Use MAICO's special tools for attaching and/or removing the eartips and
probe tubes:
Section 3.4
Section 3.5
Placing the Probe in the Ear Canal
Figure 31
5.7.4 Starting a Test (Probe Check)
IMPORTANT NOTE: To test children with Pressure Equalizer (PE) tubes, the Probe
Check needs to be omitted. Proceed as follows to disable the AutoStart function:
• Hold down the LEFT or RIGHT arrow keys for 3 seconds until the green
LED (READY) turns off.
Once the key is released, the ERO•SCAN
Probe Check
To start a test, insert the probe into the ear and select either the LEFT or RIGHT
arrow key to indicate which ear will be tested.
After the test ear is selected, the PROBE CHECK screen is shown. This screen shows if
• you have selected a proper eartip size for the patient`s ear canal volume.
• if you have placed the probe and the eartip well to obtain a seal.
See Table 9 for the various status displays on the PROBE CHECK screen.
1
8106556 Rev. 11
Operation Manual ERO•SCAN
Select an earplug that is appropriate for the patient's ear.
Apply the eartip onto the OAE probe tip
Disposables – Eartips
Disposables – Probe Tubes
Insert the eartip into the patient's first ear to test. For newborn
hearing screening do this by pulling gently down and out on the
patient's ear lobe to open the ear canal, for older patients pull
the patient's ear lobe up and back. Hold the probe and aim and
twist (gently) the eartip into the ear canal. The fit of the eartip
should be secure; not superficial. Release the earlobe. You
should not hold the OAE probe during the measurement since
this can cause acoustic noise (Figure 9).
37
®
will calibrate and test as normal.
®