Advertisement

5.7.6 Viewing Test Results
Figure 36
Figure 37
1
8106556 Rev. 11
Operation Manual ERO•SCAN
When testing is complete, a result screen (e.g. Figure 36)
appears. The result screen shows the tested ear as well as
one of the following test results:
Pass
The patient passed the screening.
No further action necessary.
Refer
The patient did not pass the screening.
Review the test result (see below) and consider
whether the test should be repeated.
Noisy Excessive noise was present during the test.
Reduce noise and repeat the test.
Section
5.1.3
No
A seal was not maintained throughout the test.
Seal
Try another eartip size and retest.
FIT
Probe placement in the ear canal is inadequate to
ERR
produce target stimulus intensities.
Try to achieve a better probe fit and retest.
NOTE: The results of the test are automatically saved in the
memory as soon as the test is complete. The results are saved
even if the device turns off or the batteries are temporarily
depleted.
Section 5.9.2
Reviewing Test Results
It is possible to review the test results in two ways:
• Review on the screen: Press
Review
(Figure 36 and e.g. Figure 37).
• Review the printout: see
Section 5.8.2
39
Noise Sources
Saving Results
to see the graph view of the test result
Interpreting Printed Results
®