Jeio tech TC-ME-025 Operating Manual page 28

Table of Contents

Advertisement

(10) Electric Circuit Breaker
In order to protect the user, sample & unit, the suitable electric circuit breaker is adopted to meet the unit's electrical
requirements and automatically cut off the electrical power in case a short circuit, or an electric leakage.
(11) Communication Port
- RS-232 Communication port is a default setting.
- RS-485 Communication port is an optional which can control
Up to 9 units at the same time.
(12) Door Limit Switch
If door is open by chance, or accidentally, it gives the signal alarm with stopping the operation of inner blower for
user's safety by using magnetic door limit switch. It also reduces the radical change of temperature or humidity inside
of chamber.
(13) Shelve
- 2 shelves are basically provided for the supporting the sample, which made of stainless steel.
- Its appearance has a fine view & clean body due to electro polishing finish.
(14) Caster
The unit can be placed on a level on the ground
Even if it's not even ground by adjustment of caster foot.
(15) Water tray
By any chance of a water leakage, or door chamber condensate
Will build up and drop into this water tray.
Just keep it clean all the time by using a dry cloth.
(16) Water Drain port
There are 2 drain condensate ports which should be connected to a silicon rubber hose over the drain barb until is
seated against the collar by tying it with a cable tie.