8. Error messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
AL
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
No pressure
An adequate pressure cannot be
in the cuff
generated in the cuff. A leak may have
AK
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
«Err 5»
Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
«LO»
Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measure-
ment.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Microlife BP A1 Easy

9. Safety, care, accuracy test and disposal

m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient's feedback.
Calling a doctor or an ambulance is advised if needed.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed. Be aware of
the risk of strangulation in case this device is supplied with
cables or tubes.
m
Contra-indications
Do not use this device if the patient's condition meets the following
contra-indications, to avoid inaccurate measurements or injuries.
 The device is not intended for measuring blood pressure in
pediatric patients of age younger than 12 years old (children,
infant, or neonates).
 Presence of significant cardiac arrhythmia during measurement
may interfere with blood pressure measurement and affect the
reliability of blood pressure readings. Consult with your doctor
about whether the device is suitable for use in this case.