Which Batteries And Which Procedure; Using Rechargeable Batteries; Error Messages; Safety, Care, Accuracy Test And Disposal - Microlife BP A80 Manual

Hide thumbs Also See for BP A80:
Table of Contents

Advertisement

Available languages

Available languages

  • ENGLISH, page 18
cannot take any further measurements and must replace the
batteries.
1. Open the battery compartment @ at the back of the instrument
by pushing inwards at the two arrows and pulling out the
battery compartment cover.
2. Replace the batteries - ensure correct polarity as shown by the
symbols in the compartment.
Which baneries and which procedure?
^
Please use 4 new, long-life 1.5V, size A A batteries.
^
Do not use batteries beyond their date of expiry.
Remove batteries, if the instrument is not going to be used
for a prolonged period.

Using rechargeable batteries

You can also operate this instrument using rechargeable batteries.
^
Please use only type «NiMH» reusable batteries!
The batteries must be removed and recharged, if the battery
symbol (battery flat) appears! They must not remain inside
the instrument, as they may become damaged (total
discharge as a result of low use of the instrument, even
when switched off).
Always remove the rechargeable batteries, if you do not
intend to use the instmment for a week or more!
The batteries can NOT be charged in the blood pressure
monitor! Recharge these batteries in an external charger
and observe the information regarding charging, care and
durability!

8. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g.
Error
Description Potential cause and remedy
<ERR1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«ERR 3»,
is displayed.
Description Potential cause and remedy
Error
<<ER R 2»
During the measurement, error signals
Error signal
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«ERR 3»
No pressure
An adequate pressure cannot be gener­
in the cuff
ated in the cuff. A leak may have
occurred. Check that cuff and pump ball
are correctly connected and that the cuff
is not fitted too loosely. Replace the
batteries if necessary. Repeat the
measurement.
«ER R 5»
Abnormal
The measuring signals are inaccurate
result
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«Н 1»
The pressure in the cuff is too high (over
Pulse or cuff
300 mmHg) OR the pulse is too high (over
pressure too
high
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«10»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
*
Please consult your doctor, if this or any other problem occurs
repeatedly
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
9. Safety. Care, Accuracy Test and Disposal

Safety and protection

This instrument may be used only for the purpose described
in this booklet. The manufacturer cannot be held liable for
damage caused by incorrect application.
This instrument comprises sensitive components and must
be treated with caution. Observe the storage and operating
conditions described in the «Technical Specifications»
section!