KLS Martin group LIMAX 60 Instructions For Use Manual page 83

Table of Contents

Advertisement

Instructions for Use
Surgical Laser Unit Limax
3
Checking required monitoring, safety and alarm devices
[ ] 3.1
Alert check (acoustic/optical signals)
[ ] 3.2
Keylock switch
[ ] 3.3
Emergency stop switch
[ ] 3.4
Interlock device
[ ] 3.5
Connection of ext. warning lamps/signs
4
Electrical safety (DIN EN 60601)
[ ] 4.1
Protective conductor resistance:
[ ] 4.2
Ground leakage current:
NOTICE
To prevent the unit from getting damaged, it is not permitted to measure the insulation
resistance (IEC 62353, section 5.3.4)
5
Measuring safety-related output parameters
[ ] 5.1
Check calibration with external power meter
(acceptable tolerance: ± 20%):
5 W / 10 W / 15 W / 30 W / 60 W
WARNING
If the unit is found to be defective in any safety-related way, it must be withdrawn from
service until the defects have been corrected.
6
Battery
[ ] 6.1
Check internal battery (U
Measures taken:
[ ] Technical Service contacted on: ......................................................................................................
[ ] Unit withdrawn from service on: ......................................................................................................
[ ] Results entered in medical device logbook on: ..............................................................................
Person performing the checks/tests:
Date :
Signature :
V. 1.0
®
60
≥ 2.8 V); replace battery (CR2032 lithium) if
batt
necessary
≤ 0.2 
≤ 0.5 mA
83