Timeouts - KLS Martin group LIMAX 60 Instructions For Use Manual

Table of Contents

Advertisement

Instructions for Use
Surgical Laser Unit Limax
6.16.6

Timeouts

> Menu timeout
The menu timeout defines for how long the rotary switch remains active after its last use.
Luminous ring white –> timeout (30 s – 5 min) –> luminous ring off
> Setting timeout
The setting timeout defines the period of time during which an activated parameter can be
changed. Changed parameter values are saved upon timeout.
Luminous ring orange / blue –> timeout (1 s – 30 s) –> luminous ring white
> Logfile timeout
The logfile timeout defines the "inactive time" period to elapse before the confirmation
query for saving the current treatment data is displayed.
V. 1.0
®
60
Setup / Timeouts
Use < BACK to exit the menu.
71