6.1.7

Keypad

NOTICE
The PILOT and LASER READY buttons are blocked during laser light emission.
All other buttons can be operated also during laser activation.
Pressing the STANDBY button switches the laser from LASER READY mode (yellow LED) to
STANDBY mode (green LED).
The rotary switch is set inactive during laser light emission. It is not possible to change
parameter settings during laser activation.
42
STANDBY button (green LED)
Pressing this button switches the laser to STANDBY (laser not ready,
green LED). While this operating state allows you to set laser
parameters, no laser light can be emitted except the pilot laser light.
The STANDBY button can also be used to stop the activated laser
during laser operation.
LASER READY button (yellow LED)
Pressing this button switches the unit to LASER READY (yellow LED),
provided a suitable laser fiber has been connected. The pilot laser is
activated automatically in this state.
PILOT button (yellow LED)
Pressing this button activates the pilot laser (aiming beam) in
STANDBY mode, provided a laser fiber has been connected.
Switching off the pilot laser in LASER READY mode switches the unit to
STANDBY mode (laser not ready for operation).
AIR button (blue LED)
Pressing this button switches the purge air flow (Flow Air) on or off. If
the current program includes parameter settings for the smoke-
clearing purge air flow, these will be automatically activated with the
program.
If no airflow parameters are stored in the current program and the AIR
function is used nonetheless, default settings will be used (see also
section 6.15.3 "Adjusting Programs", page 61).