Quick Instructions For Use Instructions For Use - KLS Martin group LIMAX 60 Instructions For Use Manual

Table of Contents

Advertisement

7

Quick Instructions for Use Instructions for Use

74
Preparation
Read user manual.
Connect unit to power supply
Connect laser connecting cable to a separate socket.
Turn on unit
Insert key in vertical position, then turn it to
horizontal position and wait until self-test has been
completed.
Select laser program
Use rotary pushbutton to set laser power, pulse type,
pulse parameters and work airflow.
Check
Inspect laser fiber for cleanness and connect it to
laser unit.
Mark laser area
Close doors leading to the treatment room.
Switch on warning lamp(s).
Put on protective goggles
Protect the patient's eyes.
Instructions for Use
Surgical Laser Unit Limax
®
60
V. 1.0

Advertisement

Table of Contents
loading

Table of Contents