Instructions for Use
Surgical Laser Unit Limax
3.8

Safety Instructions Relating to the Pilot Laser

CAUTION
Risk of irritation by pilot laser light!
The direct or reflected pilot laser light can have a glaring effect that may temporarily affect
the working ability of users.
The protective goggles provide protection only against the working laser light, not
however against the pilot laser light.
Never stare directly into the pilot laser light!
3.9

Laser Safety Officer

According to the relevant (German) regulations for the prevention of accidents (relating to
laser radiation – BGV B2), the operator of the system is required to formally appoint a Laser
Safety Officer, whose responsibilities include the following:
carry out all safety measures required;
provide instruction, to all persons involved, on required safety measures and the proper
use and operation of the unit;
mark the laser area;
check the alarm signals for proper functioning;
ensure the availability of laser protective goggle;
keep the key to the laser system in a safe plac;
ensure the proper connection of the laser after relocatio;
proper maintenance of the medical device logbook.
3.10

Users

The treating physician is responsible for the safe and proper use of the medical laser system.
He / she must check and ensure that all safety measures are duly observed and must be
familiar with laser-specific operating techniques.