Uv Lights; Work Area - Thermo Scientific 1500 A2 Series Operating Manual

Class ii, type a2 biological safety cabinet
Table of Contents

Advertisement

UV Lights

Work Area

Thermo Scientific
The optional, factory-
installed UV light is
installed at the top of the
front section of the sample
chamber.
The operating time of the
UV light is preset for 1
hour. The UV disinfection
procedure can be started by
pressing the "UV" key on
the control panel with the
window in the closed
position. If the window is
opened, the procedure is
immediately cancelled.
Note To protect from UV radiation, the UV light can only be turned on
when the window is completely closed.
The single piece work tray is provided with the cabinet. It is placed onto
the frame above the sample chamber drain pan. It can be removed by
lifting up, using the large holes in the work tray.
The working area for optimal product protection extends over the entire
width and depth of the work tray. The two armrests are positioned in the
center of the working area 8 inches (20 cm) from each other. To attach
the armrests, insert into the first perforated track of the work tray.
Warning Work safety is assured only if the armrests are used correctly!
Work tray
Perforation
track
Armrest
Separation
distance
Armrest
Holes for padded
armrests (Optional)
Figure 1-8. Work tray
Figure 1-7. Lights
s
Width
Depth
Perforation track
Section 1
Description
UV light
Chamber light
s
1500 Series A2
1-9