Section 4
Operation
Standby Mode
(continued)

Idle Mode

4-10
Model 1500 Series A2
Standby mode is entered when you close the window from within a
working session.
A Standby Mode header appears to show the unit is in standby mode,
maintaining cleanliness and containment, and ready to resume work when
you move the window back to the working position. The green "Health
Status Good" icon on the Info & Health Status Area signifies that
everything is working properly.
In the Control Panel Area, the buttons are highlighted to indicate that
functions are still on, including the blowers.
If the window is closed, the blowers will switch to reduced flow, as
indicated by the small blue circle with the "1/2" label .
Idle mode is entered when you turn the fans off, it is better for user to
make sure that the window is either in the working position or closed.
Figure 4-14 shows that it is in Idle mode with the window closed. Idle
mode is indicated by the gray "no activity" icons in the downflow and
inflow status indicator boxes.
Figure 4-14. Main Screen - Unit in Idle Mode, window closed