Clean the Drain Pan

Clean the Paper
Catch Grid
Thermo Scientific
Clean the drain pan, using a solution of tepid water and commercially
available mild dishwashing agent.
1. Remove the work tray(s) from the chamber.
2. Clean thoroughly to remove any residues and/or deposits.
3. Wipe the drain pan, using a clean cloth and plenty of clean water.
4. Discard any liquid in the drain pan. Rinse and dry thoroughly.
Note After cleaning, make sure that all cleaning product has been removed
completely from the drain pan.
5. Re-install the work tray.
The protective paper catch grids are
retained in the airduct wall by their
own tension.
1. To remove a grid section, push
the retaining tab down until the
section can be moved out from
under the inner back wall.
2. To install the grid section, first
place it on the lip of the back
wall, then push the retaining tab
down and towards the back
until the locking tabs are secured
behind the inner back wall.
Caution Do not operate the unit
without the paper catch grids
installed!