Section 2
Installation

Moving the Unit

2-4
1500 Series A2
To prevent tilting, always transport the cabinet separate from the stand,
using equipment capable of supporting the weight, even for a move within
a building.
Caution When moving the unit, lift only at the lift points shown in
Figure 2-3.
s
Caution Do not allow the weight of the cabinet to rest on the drain pan!
Warning The weight of the window is balanced by the counterweight on
the back of the cabinet. Do not move the unit unless the counterweight
has been locked in place with the shipping screws (see Figure 3-2).
Figure 2-3. Lift points
s
s