7.12 Alarms, Warnings and Errors

The SelfCheck Assistant of your Proline Kryomat monitors more than 50 device parameters and triggers
alarms, warnings or errors as appropriate.
All warnings and alarms are shown on the Command remote control in plain text. Errors are shown in
plain text on the Command remote control, also, in an error list.
Alarms:
Alarms are safety relevant. Pump, heater and refrigerating unit will be shut off.
Warnings: Warnings normally are not safety relevant. The device continues to operate.
Errors:
If an error occurs, the pump, heater and refrigerating unit switch off automatically.
Switch of the unit at the rotary mains switch. If the error is always present after switching on
the device, please give information to the LAUDA Service for Constant Temperature
Equipment ( 9.5).
Find cause of alarm or warning and rectify where necessary. Then press
order to remove the alarm message. Warning messages can be removed either on the Master keyboard
with
or on the Command board with
Warnings may be ignored by pressing
Softkey on the Command remote control. Warnings will not be repeated periodically.
7.12.1

Overtemperature protection and checking

The units are designed for operation with non-flammable and flammable liquids to
DIN EN 61010-2-010.
YAUE0007 / 10/2/2018
.
or
on the Master keyboard or by activating the Screen
Starting up
on the Master keyboard in
W303