Lauda RP 3050 C Operating Instructions Manual page 43

Hide thumbs Also See for RP 3050 C:
Table of Contents

Advertisement

Proline Kryomats
Bath temperature
02%32
Overtemp. cut-off
90
Level alarm
LEUEL
Command
English
Deutsch
YAUE0007 / 10/2/2018
 The momentary bath temperature is displayed,
 the pump starts provided "Standby" or "Manual start" ( 7.6.1) has
°C
not been programmed,
 all values are accepted which were active before switch-off.
Check or set overtemperature cut-off point:
 The switching point is shown in the LED display on pressing the key
°C
 Change overtemperature cut-off ( 7.12.1). Overtemperature
protection and checking on page 89.
 If necessary, top up heat transfer liquid, this has been pumped out
by filling the external load.
 Display for
little liquid.
°C
 Red LED
 Find cause of fault and, where necessary, top up missing heat
transfer liquid ( 6.3).
 Press the Enter key.
 Also, press the key if unit has been switched off in the fault state.
 No release is possible on Command remote control!
Français
Español
.
LEUEL
(low level) appears when the bath has too
above the fault triangle
Language
 If the Command remote control is being
switched on for the first time, the
illustrated window appears automatically,
enabling you to select the dialog language
with the appropriate soft key.
Starting up
flashes.