Restoring Works Settings - Lauda RP 3050 C Operating Instructions Manual

Hide thumbs Also See for RP 3050 C:
Table of Contents

Advertisement

7.6.4

Restoring works settings

All works settings, apart from the control parameters and the sensor calibration, are restored.
Command
All modules
Master
Command
Cool
Pump
Menu
Confirm input!
Enter key: Continue
Escape key: Cancel
Pump
Menu
64
all default
only control par. int
only control par. ext
only miscellaneous
End
T
T
set
End
T
T
set
Starting up
Works settings
 Open the device parameter menu via
the soft key
 With the cursor keys continue to
Settings
 The window shown opposite appears.
Master and then
int.
is shown as a possible choice.
There are however various possibilities,
which can be selected with
 Under
fix
and all connected modules are reset to
the works setting with
 Under
between:
all default
are reset,
only control para. int.
internal control parameters,
only control para. ext.
external,
only miscellaneous
setpoint, pump level, maximum
current consumption, control to
internal and auto start to "Auto".
 Under
are reset with
 Confirm selection with
 Confirm the control dialog shown on the
left with
 Return to measurement window with
fix
End or
Proline Kryomats
Menu .
Works settings
only control par.
All modules
Master, Command
all default
Master
you have the choice
, then all Master settings
for the
similar for
which resets
Command
all command settings
All default
.
.
or cancel with
.
YAUE0007 / 10/2/2018
.
or
:
.
.