Low-Level Alarm And Low-Level Checking - Lauda RP 3050 C Operating Instructions Manual

Hide thumbs Also See for RP 3050 C:
Table of Contents

Advertisement

Proline Kryomats
Command
7.12.2

Low-level alarm and low-level checking


Level alarm
LEUEL
Command
YAUE0007 / 10/2/2018
 Slowly lower T
 Cut-off at the bath temperature should occur.
 Step 1 – 2 (see above) must follow.
 Set the overtemperature cut-off higher than the bath temperature again
and wait until
 Unlock with the
Unlocking is not possible on the Command control elemant!
Overtemperature alarm!
Overtemperature alarm! is shown in the display and signifies that
unlocking is only possible on the Master control panel.
 If the liquid level falls so far that the heating element is no longer
completely covered with liquid, an alarm is initiated:
1. The alarm sounds as a dual-tone signal.
2. Display for
contains too little liquid.
°C
3. The red LED
 Heater switches off on both poles,
 Varioflex pump and refrigerating unit are switched off.
 Find the cause of the fault and, where necessary, top up for
missing liquid ( 6.2 and 6.3).
 Press the Enter key.
 Also, press this key if the unit has been switched off in the fault
state.
 Checking the safety system at regular intervals by lowering the bath level. To do
this, fit a hose to the drain point and slowly run off the heat transfer liquid into a
suitable container via the drain-point tap at the front.
 Step 1 – 2 must follow.
 With this test, the bath temperature must not be below 0 °C or above 50 °C,
otherwise there is a risk of injury!
 If irregularities arise during the checking of the safety devices, switch off the unit
immediately and pull out the mains plug.
 Have the equipment checked by LAUDA Service for Constant Temperature
Equipment.
Low-level alarm!
Low-level alarm! is shown in the display and signifies that
unlocking is only possible on the Master control panel.
, as described above.
max
tEMNP
appears in the display.
key.
LEUEL
(low level) is shown when the bath
above the fault triangle
Starting up
flashes.