Lauda RP 3050 C Operating Instructions Manual page 105

Hide thumbs Also See for RP 3050 C:
Table of Contents

Advertisement

Proline Kryomats
Command
IN_SP_00
IN_SP_01
IN_SP_02
IN_SP_03
IN_SP_04
IN_SP_05
IN_PAR_00
IN_PAR_01
IN_PAR_02
IN_PAR_03
IN_PAR_04
IN_PAR_05
IN_PAR_06
IN_PAR_07
IN_PAR_09
IN_PAR_10
IN_PAR_14
IN_PAR_15
IN_DI_01
IN_DI_02
IN_DI_03
IN_DO_01
IN_DO_02
IN_DO_03
IN_MODE_00
IN_MODE_01
IN_MODE_02
IN_MODE_03
IN_MODE_04
TYPE
VERSION_R
VERSION_S
VERSION_B
VERSION_T
VERSION_A
VERSION_V
VERSION_D
VERSION_M_0
VERSION_M_1
VERSION_M_3
VERSION_M_4
VERSION_M_5
YAUE0007 / 10/2/2018
Read temperature setpoint.
Read current pump power level.
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC).
Read current overtemperature switch-off point.
Read current outflow temperature limit TiH.
Read current outflow temperature limit TiL.
Read control parameter Xp.
Read control parameter Tn (181 = OFF).
Read control paramete Tv.
Read control parameter Td.
Read control parameter KpE.
Read control parameter TnE (980 = OFF).
Read control parameter TvE (0 = OFF).
Read control parameter TdE.
Read value of correction limitation
Read control parameter XpF.
Interrogation of the setpoint offset.
Read control parameter PropE.
State of contact input 1: 0 = open/ 1 = closed.
State of contact input 2: 0 = open/ 1 = closed.
State of contact input 3: 0 = open/ 1 = closed.
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed.
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed.
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed.
Keyboard Master: 0 = free / 1 = locked. .
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analogue / 3 = ext. Serial.
Standby operation: 0 = Device ON / 1 = Device OFF.
Keyboard Command remote control: 0 = free / 1 = locked.
Setpoint offset source: 0=normal/1=ext.Pt/2=ext.analogue/3=ext.serial.
Read equipment type (response e.g. RP 3050).
Read software type of control system.
Read software type of protection system.
Read software type of Command.
Read software type of cooling system.
Read software type of analogue module.
Read software type of RS-232/485 module.
Read software type of digital module.
Read software type of solenoid valve (cooling water).
Read software type of solenoid valve (automatic refill).
Read software type of solenoid valve (shut-off valve 1)
Read software type of solenoid valve (shut off valve 2).
Read software type of high temperature cooler.
Interface modules
Explanation
105