5.6.1.5 [Off]
Turn off the selected markers.
5.6.1.6 [All Off]
Turns off all the opened markers and the related functions. The
marker won't show again.
5.6.1.7 [Marker Table On Off]
Turns on or off the display of the marker table.
5.6.2【MarkerFctn ►】right softkey
Key access:
[Marker] bottom softkey → [MarkerFctn ►] right softkey
Accesses the marker function setting menu.
5.6.2.1 [Function Off]
Turn off marker measurement function.
5.6.2.2 [NdB On Off]
Enables the N dB bandwidth measurement or sets the value of N.
The N dB bandwidth denotes the frequency difference between
points that are located on both sides of the current marker while
the amplitude falls off (N<0) or rises (N>0) N dB separately,
Key points:
When the measurement starts, the analyzer will search the
two points which are located at both sides of the current point
and are N dB amplitudes smaller or greater than the current
point, and display the frequency difference between the two
points.
You can use the numeric keys, control or direction keys to
modify the value of N, 3 dB at default.