The Probe Compensation - Jinhan JDS3022E Manual

Table of Contents

Advertisement

1.5 the probe
1. The safety of the probe
Probe body surrounding protection Settings can protect the fingers to prevent electric shock.
Prior to any measurement, connect the probe to the oscilloscope and grounding terminal.
2. The probe compensation (see the probe specification)
In the first connect the probe to any input channel, this adjustment is required, make the
probe to the input channel. Without compensation correction probe will lead to a measurement
error or mistakes. If the adjustment probe compensation, please click the following steps:
(1) the channel attenuation is set to 10 x, will probe options menu to probe on the switch is
set to 10 x, and the oscilloscope probe connected to channel 1. Such as using the probe hook
head, should ensure that contact with the probe and reliable.
(2) the probe end connected to the signal generator output connector, clamp and signal
generator is connected to ground wire connector, according to channel, and then press the
"AUTO" button (automatically).
(3) Check shows the shape of the waveform. As shown in figure 1-5.
23