3.5 Attaching Condensate Drain

In case of the liquid leakage caused by spilled reagent or cracked flask inside of the chamber, the
liquid is designed to be drained out through the drain of the unit. The drain barb is made of chromium
plated copper and it is easy to connect 6mm ID soft-walled tubing to the condensate drain.

3.6 Attaching/Detaching External Refrigeration (ISF-7100, ISF-7200 Only)

Many laboratories are built with central chilling systems. With this in mind Jeio Tech has developed
the ISF-7100/7200 so that it can be connected to external refrigeration systems. To connect the unit
to an external refrigeration source use Ø 8mm OD hard-walled tubing, such as nylon or polyethylene.
Note: The ISF-7100/7200 controller does not have the ability to communicate with the external
refrigeration source. Use the following procedure to connect/disconnect the tubing to the unit. Use the
following procedure to connect/disconnect the tubing to the unit.
(1) Inserting tubing into fitting
Hold the tubing firmly and push it tightly into the fitting. Run the tubing to the external refrigeration
source.