Once the setting time is changed during the timer operating, original set time is changed. The
accumulated time is applied to the new time setting and the remaining time is calculated and
displayed. If you want whole new time setting, off the timer and set the timer again.
Timer setting is available for hour:minute. Touch the part you want to change and put the numbers
on it.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
On the timer set display, you can check the present chamber temperature and rpm. (right upper
side of the display)
Below 1 minutes is displayed as"mmm:ss".
Timer is not counted during the door open.

4.2.4 Alarm and Stop by force

The unit senses the problem on the unit and warn to the user. The warning is visual-audial effect. The
alarm is divided into warning and fault.
4.2.4.1 Warnings
Warning generates visual-audial alarm but the unit is operating. Once the warning is generated, Silent
button is appeared on the main display. Once the silent button is touched, next 15 minutes there is no
alarm. If the problem is on-going after 15 minutes, the alarm is generated again.