4.4

System Settings

Set the main variable that affects on the unit. Please read the manual carefully before setting the system.
Touch the System setting button on the home display, below clauses are arranged.
1
Control Deviation
Limit
2
External
Start/Stop Key
3

Calibration

4

Auto Tuning

5
Operating
Parameters
6

Reset

4.4.1 Control Deviation Limit

Set the deviation limit between setting value and actual value of the temperature and rpm. The time of
out of deviation limit range can be set either.
1
Temperature
Deviation Limit
2
Temperature
Deviation Alarm
Delay
3
RPM
Deviation Limit
4
RPM Deviation
Alarm Delay

4.4.2 External Start/Stop Key

External Start/Stop button is on the control penal as membrane button. It generates start
and stop. Set the applying range of it.
1
Only Shaker
2
Only Incubator
3
Shaker Incubator
Set the deviation limit range. Temperature setting and rpm setting are
available.
Start/Stop membrane button is located on the control panel. Set about
this key
Calibration on the temperature sensor and rpm to the user's reference.
Calibration on the LCD touch coordinate is available.
Temperature PID parameter update. Optimum control on the main
operating temperature is available.
Set the operating and error check
Set the parameter as initial factory value.
Set the deviation limit between setting value and actual value. Set on the
key pad.
Once the deviation is out of range for alarm delay period, the alarm is
generated. Once the actual temp reached to the set temp, the alarm is
disappeared after alarm delay period check.
Set the deviation limit between setting value and actual value. Set on the
key pad.
Once the deviation is out of range for alarm delay period, the alarm is
generated. Once the actual rpm reached to the set rpm, the alarm is
disappeared after alarm delay period check.
The button only works for shaker's start/stop.
The button only works for incubator's start/stop.
The button only works for shaker and incubator's start/stop.
28