Troubleshooting; Stop During The Operating - Jeio tech ISF-7100 Operating Manual

Shaking incubator
Table of Contents

Advertisement

6 Troubleshooting

6.1 Stop during the operating

Display
Message
Temperature
Analog over temperature regulator is
Limit
Vibration is generating and the
Load
normal shaking is not available. To
Unbalance
prevent the shaking system, stop the
Door is opened over 5 minutes
Door Open
Temperature sensor is faulty so the
Sensor Fault
Over current is generated so the
Over Current
Description
active.
shaking.
during the operating.
operating is stopped.
operating is stopped.
Check the set on the analog over
temperature regulator and the operating.
Restart after temperature is inside of the
Check the unbalanced sample setting,
overload, un-flat ground.
Close the door and touch the start button
and restart the unit.
Contact the manufacturer or sales point.
Check the overload, obstacle in the shaking
38
Solution
range.
system.