Jeio tech ISF-7100 Operating Manual page 35

Shaking incubator
Table of Contents

Advertisement

Shaker and incubator are both available for the external membrane button. Once one of the
shaker/incubator is on the generating and touch the button, the stopped part is operated. Touch
the button again, both of mode are stopped.
4.4.3 Calibration
The temperature sensor and rpm reached to the user's reference measure value, accurate experiment
circumstance can be made. Calibration for coordinate is available.
4.4.3.1 Temperature Calibration
Calibration for temperature sensor of main controller. The unit provides three points calibration.
Each point of three shows the uncalibrated and calibrated status. Touch the point you want to change.
Touch the one point of Calibration 1,2,3. Put the external sensor on the chamber and close the door.
Touch the temperature on Set Temp. and put the desired temperature on it and operate the unit. Wait until
the actual temp reaches to the setting temp. Once the actual temp reaches to the setting temp, touch the
calibrated temp and put the temperature from the external sensor. Touch confirm for the saving. The unit
stops the operating. The unit read the temp as the external sensor's value.
29