Another Problem And Solution - Jeio tech ISF-7100 Operating Manual

Shaking incubator
Table of Contents

Advertisement

6.2 Another problem and solution

Trouble
The equipment is not
on
Power on but
controller does not
operated
NO The temperature of
incubator rise to setting
value
Shaking trouble
Check the power plug connection to the socket
Check if the circuit breaker is operated
When the fuse is short circuited, replace new one as enclosed.
Check the power failure
If the problem is not fixed with the above method, please request A/S
Check the power plug connection to the socket
Check the panel board circuit breaker and reset it then turn the
machine on
Request A/S if Display board or Main PCB has problem
Request A/S if Power switch has problem.
① Check the Temperature Control ON indicator is ON
② Check outdoor temperature is too high or low and maintain the right
room temperature.
③ Request A/S after check the difference between actual temperature
and display temperature.
④ Check the door switch irregularity
⑤ Request A/S if the heater is out of order
⑥ Request A/S if the freezer is out of order.
⑦ Check air filter condition and clean it according to 5.3
⑧ Auto Tuning if the temperature is unstable according to 4.4.4.
⑨ If the problem is not fixed with the above method, please request A/S.
① If there is extremely intense vibration of the machine than usual during
operation, check the horizontality. If you need, please adjust
horizontality according 3.4.2
② If there is extremely intense vibration of the machine and shaking
39
Probable Cause & Solution

Advertisement

Table of Contents
loading

This manual is also suitable for:

Isf-7200Isf-7100rIsf-7200r