When using the 3-Point Known Load Calibration type, the following menu items
are available:
"
": point 1 of 3, usually 0% of the transducer's full-scale capac-
KNOWN POINT 1/3
ity.
"
": point 2 of 3, usually 50% of the transducer's full-scale capac-
KNOWN POINT 2/3
ity.
"
": point 3 of 3, usually 100% of the transducer's full-scale
KNOWN POINT 3/3
capacity.
When using the 5-Point Known Load Calibration type, the following menu items
are available:
"
": point 1 of 5, usually 0% of the transducer's full-scale capac-
KNOWN POINT 1/5
ity.
"
": point 2 of 5, usually 25% of the transducer's full-scale capac-
KNOWN POINT 2/5
ity.
"
": point 3 of 5, usually 50% of the transducer's full-scale capac-
KNOWN POINT 3/5
ity.
"
": point 4 of 5, usually 75% of the transducer's full-scale capac-
KNOWN POINT 4/5
ity.
"
": point 5 of 5, usually 100% of the transducer's full-scale
KNOWN POINT 5/5
capacity.
NOTICE
To insure both correct operation of the transducer and application of the load,
the instrument expects the voltage applied at each known-load point to be
increasing. For example, the load applied at Known-Load Point 2/2 must
cause the transducer to produce a more positive voltage than at Known-Load
Point 1/2.
These menu items are only available with the Known-Load Calibration types.
SC Series Instruction Manual - Model SC500
page 63