Tare] Button; Indicator Lights - Honeywell SC500 Manual

Table of Contents

Advertisement

Shunt Calibration Check

3.3.5 [TARE] button

3.3.6 Indicator Lights

Model SC500
button cannot be used to change which channel the lower line is monitoring. That
is selected with the "
DISPLAY MENU
If the [CHANNEL] button is held down for more than 3 seconds when the display
is monitoring the Input Channel (Channel 01), the Input Channel's shunt calibra-
tion function (if the channel has shunt calibration available) will be engaged.
To reset the channel's display to zero, press the [TARE] button while in RUN
mode. To restore the tare value, press the [TARE] button again. The "Tare" indica-
tor will illuminate when tare is on.
To tare the Input Channel regardless of which channel is being monitored by the
display, use the AUX1 or AUX2 control pins on the Input Channel connector. See
the particular chapter regarding the type of channel for details.
A "
" symbol on the right side of a channel's display indicates tare is off for that
"
channel; a "
" symbol indicates tare is on.
Ë
If the display is configured with the "
"
", then the lower line will monitor the status of the Limits. A "
LOWER=LIMITS
indicates a limit is deactivated; a "
example, an instrument with two limits might have the lower line of its display look
like this:
which indicates that Limit 1 is deactivated and Limit 2 is activated. See "DISPLAY
MODE Menu Item" on page 43.
NOTICE
These indicators may be overridden by the operation of a Mathematics Virtual
channel in custom applications identified at time of order.
SC Series Instruction Manual - Model SC500
->
" menu item.
LOWER CHANNEL
->
DISPLAY MENU
DISPLAY MODE
" symbol indicates a limit is activated. For
Ë
L1" L2Ë
" menu item as
" symbol
"
page 21