11.4 Specifications

page 76
008-0671-00
TRANSDUCER INPUT
Transducer type
Excitation Voltage
Transducer full-scale output
Amplifier Gain Selection
Calibration Type
A/D Converter
Low-pass filter
Resolution
Frequency Response
AUXILLIARY INPUTS
Quantity
Response Time
Field-Selectable Functions
DIGITAL-TO-ANALOG OUTPUT
Availibility
Output voltage range
Source
Isolation
Resolution
Frequency Response
full-bridge, AC/AC-LVDT
3 VRMS @ 5kHz,
short circuit protected
.1 to 15 VRMS @ 3 VRMS excitation
automatic
2-, 3- or 5- point known displacement
24-bit Sigma-Delta
digital, 24-tap FIR
see page 77
see page 77
Type
momentary contact closure
< 5ms
tare on, tare off,
peak/valley clear, peak/valley hold,
track hold
with Option 58i
5, ±5, 10 or ±10 VDC (field selectable)
any channel's track, peak or valley value
500V
14 bits
same as input if driven by the same chan-
nel's tracking data
2