10.2.2 Strain-Gage
Transducers with
Buffered Shunt Cal
page 54
008-0671-00
Some strain-gage transducers with mV/V output include a "buffered shunt calibration"
feature. Connecting the transducer's Shunt Cal pin to its (-)Excitation pin, activates its
shunt calibration function. Transducers with this feature include:
Model FP2000 pressure transducer with Options 2u and 3d
Figure 10-3: Transducer w/Buffered Shunt Cal Connection to Strain-Gage Input Channel
To use the transducer's buffered shunt cal feature, the instrument must be configured
to connect the SHUNT1 and SHUNT2 pins of the Input Channel connector when the
shunt cal feature is needed. If transducer and instrument were purchased together,
the instrument was configured for the correct Shunt Cal Application configuration.
Table 4-8: Shunt Cal Application Configuration
Does transducer
Instrument's internal Shunt
have "buffered
Cal Resistor (see "Figure 10-
shunt cal" feature?
no
yes
Figure 10-4: Location of Shunt Cal Application Jumpers
2:" on page 53)
installed
removed
Jumper Block J24
pins
pins
pins
1 & 2
3 & 4
5 & 6
open
open
closed
closed
closed
open